AT1321 Spectrometric Personal Radiation Detector

AT1321 Spectrometric Personal Radiation Detector (SPRD)

– Search and detect gamma radiation sources with identification of radionuclide composition
– Measure gamma radiation ambient dose equivalent rate

 

  • Highly-sensitive to gamma radiation and compact
  • Spectrum analysis and radionuclide identification without PC
  • Integrated GPS module
  • Sound, light and vibration notification
  • “GARM” application software for further data processing and analysis in expert mode and radiological mapping

 

Request a quote

Specification

AT1321
Detectors Scintillation NaI(Tl), Ø25×40 mm;
Geiger-Muller counter tube
Energy range 20 keV – 3 MeV
Detectable activity of 137Cs source, located at the distance of 15 cm in a time not longer than 2 s (50±10) kBq
Typical resolution at 662 keV (137Cs) 8.5%
Measurement range of ambient dose equivalent rate 30 nSv/h – 100 mSv/h
Limits of tolerable intrinsic relative error
±20%
Typical sensitivity to 137Cs gamma radiation 425 cps/(µSv·h-1)
Energy dependence relative to 662 keV (137Cs) ±20% (50 keV – 3 MeV)
Response time for dose rate change from 0.1 to 1 µSv/h <2 s
Protection class IP54
Overall dimensions, weight 145x100x50 mm, 0.7 kg

Design and specifications are subject to change without notice.
For details please refer to the datasheet.

Downloads

You have not viewed any product yet!