Specification
AT6130 Available models |
||
AT6130 | AT6130-412152.005-06 | Gamma channel – Sv, beta channel – min-1/cm-2 |
AT6130-412152.005-12 | Gamma channel – Sv, beta channel – CPS | |
AT6130-412152.005-14 | Gamma channel – rem, beta channel – CPS | |
AT6130-412152.005-16 | Gamma channel – rem, beta channel – min-1/cm-2 | |
AT6130-412152.005-22 | Gamma channel – Sv, beta channel – Bq/cm2 | |
AT6130-412152.005-24 | Gamma channel – rem, beta channel – Bq/cm2 | |
AT6130A | AT6130A-412152.005-08 | Gamma channel – Sv |
AT6130A-412152.005-18 | Gamma channel – rem | |
AT6130D | AT6130D-412152.005-10 | Gamma channel – Sv |
AT6130D-412152.005-20 | Gamma channel – rem |
AT6130, AT6130A, AT6130D Radiation Monitors | |
Measurement range of ambient dose equivalent rate | 0.1 µSv/h – 10 mSv/h (AT6130, А) 0.1 µSv/h – 100 mSv/h (AT6130D) |
Measurement range of ambient dose equivalent | 0.1 µSv – 100 mSv (AT6130, А) 0.1 µSv – 1 Sv (AT6130D) |
Measurement range of beta particle flux density | 10 – 104 particle·min-1·cm-2 (AT6130) |
Limits of tolerable intrinsic relative error | ±20% |
Energy range of X-ray and gamma radiation | 20 keV – 3 MeV (AT6130) 50 keV – 3 MeV (AT6130A, D) |
Energy range of beta radiation | 155 keV – 3.5 MeV (AT6130) |
Energy dependence relative to 662 keV (137Cs) | ±30% |
Time of continuous operation | ≥500 h |
Protection class | IP57 |
Drop protection | ≤1.5 m height |
Overall dimensions, weight | 110x60x38 mm, 0.25 kg |